Sieve calibration
McManus, D.A. (1963). Sieve calibration. J. Sediment. Petrol. 33(4): 953-954
In: Journal of Sedimentary Petrology. Society of Economic Paleontologists and Mineralogists (SEPM): Tulsa. ISSN 0022-4472, meer
Peer reviewed article  

Beschikbaar in  Auteur 

Auteur  Top 
  • McManus, D.A.

Abstract
    The average aperture size of a sieve may vary 0.08 phi-unit from the stated aperture size because of manufacturing tolerances. For sediment studies requiring greater than routine precision this possible error can be eliminated by calibrating the sieves with samples of glass spheres especially prepared for sieve calibration by the National Bureau of Standards. The routine particle size analysis of sediments, however, does not require the use of calibrated sieves.