Het OWA, het open archief van het Waterbouwkundig Laboratorium heeft tot doel alle vrij toegankelijke onderzoeksresultaten van dit instituut in digitale vorm aan te bieden. Op die manier wil het de zichtbaarheid, verspreiding en gebruik van deze onderzoeksresultaten, alsook de wetenschappelijke communicatie maximaal bevorderen.
Dit archief wordt uitgebouwd en beheerd volgens de principes van de Open Access Movement, en het daaruit ontstane Open Archives Initiative.
Basisinformatie over ‘Open Access to scholarly information'.
one publication added to basket [221375] |
Hf and Lu isotopic reference values for the zircon standard 91500 by MC-ICP-MS
In: Chemical Geology. Elsevier: New York; London; Amsterdam. ISSN 0009-2541; e-ISSN 1872-6836, meer
| |
Author keywords |
Hf isotopes; Zircon; MC-ICP-MS; Zircon standard 91500; Lu isotopes |
Auteurs | | Top |
- Goolaerts, A.
- Mattielli, N., meer
- de Jong, J., meer
|
|
|
Abstract |
The Hf and Lu isotopic ratios of the zircon standard 91500 have been redefined by multiple collector inductively coupled plasma mass spectrometry (MC-ICP-MS). The chemical protocol consists of a high-pressure and high-temperature acid dissolution of zircon followed by optimized two-step ion-exchange chromatography with an overall recovery better than 94%. Instrumental calibration involved an extensive characterization of standard solutions for Hf [JMC-475 250 ppb: 176Hf/177Hf=0.282161±16 (2 S.D.)] and Lu [JMC standard 50 ppb: 175Lu/176Lu=37.72±2 (2 S.D.)]. A total of 59 Hf isotopic measurements on the zircon standard 91500, including four independent dissolutions and purifications, were carried out on two Nu Plasma MC-ICP-MS instruments. The average 176Hf/177Hf is 0.282302±8 (2 S.D.) and the two MC-ICP-MS instruments give overlapping results within error. These high precision data [external error: 27 ppm (2 S.D.) for 59 measurements and 10 ppm for a single day; internal error: 10 ppm (2 S.D.)] are reproducible between duplicates and underscore the stability of the analytical protocol. The Lu isotopic composition of the zircon standard 91500 has also been redefined: 175Lu/176Lu=37.74±8 (2 S.D.). With the growing use of in situ measurement techniques using MC-ICP-MS technology, such precise reference values will be important for the calibration of laser ablation analyses in the future. |
IMIS is ontwikkeld en wordt gehost door het VLIZ.